Applications
(1) Environmental and Emergency Response Monitoring – On-site rapid monitoring of chemical reagent leaks and acidic gas releases during accidental spills and emergency incidents.
(2) Industrial Process Monitoring – Low-emission monitoring across multiple industries, including chemical plants, power stations, steel mills, and cement plants, with simultaneous measurement of multiple gas components such as SO₂, NOₓ, and others.
(3) Emission Testing – Flue gas emission monitoring for municipal solid waste incineration, medical waste incineration, hazardous waste incineration, as well as engine exhaust gas analysis.
(4) Portable Routine Analysis – On-site rapid qualitative screening and quantitative analysis for general field inspection purposes.
Performance Features
(1) Solid Corner-Cube Pendulum Interferometer– Compact in size, lightweight in design, and highly resistant to mechanical shock, ensuring long-term interferometric stability.
(2) Moisture-Proof ZnSe Beam Splitter– Equipped with a moisture-resistant ZnSe beam splitter specifically designed for demanding field environments.
(3) Multi-Sealed Moisture-Proof Design– The interferometer features a multi-sealed moisture-proof structural design that effectively protects the optical path system from interference caused by high temperature, high humidity, and chemical corrosion in various aspects.
(4) Oxygen Measurement Module– The gas cell integrates a zirconia (ZrO₂) oxygen sensor, enabling simultaneous measurement of IR gas components and oxygen concentration.
(5) Analysis Software– Capable of continuous automated operation and real-time measurement of a wide range of organic and inorganic gas components. Additionally, it supports seamless transfer of analytical models through spectral data comparison between different instruments.
(6) Chemometrics Software– Developed with a user-centric approach, this chemometrics software features an intuitive interactive interface and user-friendly operation for efficient one-click intelligent modeling. It is compatible with multiple spectral data formats and offers comprehensive functionalities including dataset partitioning, preprocessing methods, wavelength selection, and model transfer, supporting both qualitative and quantitative multi-component analysis.
(7) Model Prediction Toolkit– The model prediction toolkit supports encryption and third-party development, enabling direct prediction results by inputting only the predictive model and corresponding spectra.
Specifications
| Spectrometer | |
|
Wavenumber Range |
800cm-1-4400cm-1 |
|
Scan Speed |
Supports fast scanning (multiple speed levels selectable) |
|
Light Source |
SiC |
|
Beam Splitter |
ZnSe |
|
Detector |
Electrically Cooled MCT |
| Gas Cell | |
|
Type |
Multiple-pass reflection, fixed path length of 5 m / 10 m |
|
Mirror |
Gold-plated with protective coating |
| Performance Specifications | |
|
Limit of Detection (LOD) |
≤0.1~1ppm |
|
Linearity Error |
≤ ±5% of Full Scale (F.S.) |
|
Ambient Temperature Effect |
≤ ±5% of Full Scale (F.S.) |
|
Water Vapor and CO₂ Interference |
≤ ±4% of Full Scale (F.S.) |